Please use this identifier to cite or link to this item: http://repositorio.unicamp.br/jspui/handle/REPOSIP/99488
Type: Artigo de periódico
Title: Fine Needle Aspiration Biopsy Of Solid Tumours By Auto-vacuum System: A Study In Rats
Author: De Freitas Jr. R.
Giraldo P.C.
Rettori O.
Vieira-Matos A.N.
Tambascia J.K.
Abstract: A fine needle aspiration biopsy (FNAB)) device was designed to utilize a steel spring located between the syringe and its plunger which exerts a constant negative pressure (auto-vacuum). In 43 rats inoculated with breast tumor cells the technique was compared with the standard procedure of FNAB (21 Sp1 and 22 Walker 256 carcinosarcoma). Malignant cells were cytologically confirmed in all cases. Sufficient material was obtained in 95% of FNAB by auto-vacuum system versus 86% by standard technique. The superior results obtained with the auto-vacuum system was seen for both soft and hard tumours.
Editor: 
Rights: fechado
Identifier DOI: 
Address: http://www.scopus.com/inward/record.url?eid=2-s2.0-0027073119&partnerID=40&md5=73feb5b028836b461c39ab1cf1618850
Date Issue: 1992
Appears in Collections:Unicamp - Artigos e Outros Documentos

Files in This Item:
There are no files associated with this item.


Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.