Please use this identifier to cite or link to this item: http://repositorio.unicamp.br/jspui/handle/REPOSIP/94920
Type: Artigo de periódico
Title: Clinical Study Of The Sensitivity And Dynamic Range Of Three Digital Systems, E-speed Film And Digitized Film.
Author: Boscolo F.N.
Oliveira A.E.
Almeida S.M.
Haiter C.F.
Haiter Neto F.
Abstract: The objective of the present study was to clinically evaluate the sensitivity and dynamic range of three digital systems (Sens-A-Ray, CDR, Digora), digitized film and E-speed film. Five objects were submitted to three different kilovoltages and seven exposure times. In order to evaluate the relationship between exposure time and dosage, measurements were made in a dental x-ray unit with an ion chamber, X-ray timer and kVp meter. For each system, 105 radiographs were taken, totalling 525 images, that were evaluated by six trained observers by means of scoring from 0 to 4. The scores attributed to the images for each of the systems were submitted to analysis of variance and Turkey's test. The results showed statistically significant differences (p < 0.01) with the storage phosphor system producing the wider dynamic range, followed by the digitized film. The CDR system showed the greater sensitivity, followed by the Digora system.
Editor: 
Rights: aberto
Identifier DOI: 
Address: http://www.scopus.com/inward/record.url?eid=2-s2.0-0035238301&partnerID=40&md5=682ee64f92c13bcfddadc4cdd1951824
Date Issue: 2001
Appears in Collections:Unicamp - Artigos e Outros Documentos

Files in This Item:
There are no files associated with this item.