Please use this identifier to cite or link to this item: http://repositorio.unicamp.br/jspui/handle/REPOSIP/87983
Type: Artigo de periódico
Title: Fractographic Analysis Of 2.0-mm Plates With A Screw Locking System In Simulated Fractures Of The Mandibular Body
Author: De Medeiros R.C.
Lauria De Moura A.
Rodrigues D.C.
Menezes Mendes M.B.
Sawazaki R.
Fernandes Moreira R.W.
Abstract: Purpose The purpose of the present study was to analyze the fractured plates from 2 brands of 2.0-mm locking fixation systems submitted to axial linear load testing. Materials and Methods Four aluminum hemimandibles with linear sectioning to simulate a mandibular body fracture were used as a substrate and fixed with 2 fixation techniques from 2 national brands: Tóride and Traumec. The techniques were as follows: one 4-hole plate, with four 6-mm screws in the tension zone, and one 4-hole plate, with four 10-mm screws in the compression zone; and one 4-hole plate, with four 6-mm holes in the neutral zone. The hemimandibles were submitted to vertical linear load tests using an Instron 4411 mechanical test machine. The system was submitted to the test until complete failure had occurred. Next, a topographic analysis of the surface of the plates was performed using a stereomicroscope and an electronic scanning microscope. The samples were evaluated using different magnifications, and images were obtained. Results The surface of the fracture analyzed in scanning electron microscopy demonstrated a ductile-type fracture, usually found in the traction test bodies of ductile materials, such as titanium. No evidence of failure was observed in any fracture surface from a change in the structure or composition of the material. Conclusions The plates were fractured by a ductile rupture mechanism, as expected, suggesting that the manufacturing of the national brand name plates used in the present study has been under adequate quality control, with no structural changes produced by the manufacturing process that could compromise their function. © 2014 American Association of Oral and Maxillofacial Surgeons.
Editor: W.B. Saunders
Rights: fechado
Identifier DOI: 10.1016/j.joms.2014.01.003
Address: http://www.scopus.com/inward/record.url?eid=2-s2.0-84901244823&partnerID=40&md5=6f5c0911438565b536ef214966fe58a4
Date Issue: 2014
Appears in Collections:Unicamp - Artigos e Outros Documentos

Files in This Item:
There are no files associated with this item.


Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.