Please use this identifier to cite or link to this item: http://repositorio.unicamp.br/jspui/handle/REPOSIP/73125
Type: Artigo de periódico
Title: Validation of diffuse correlation spectroscopic measurement of cerebral blood flow using phase-encoded velocity mapping magnetic resonance imaging
Author: Buckley, EM
Hance, D
Pawlowski, T
Lynch, J
Wilson, FB
Mesquita, RC
Durduran, T
Diaz, LK
Putt, ME
Licht, DJ
Fogel, MA
Yodh, AG
Abstract: Diffuse correlation spectroscopy (DCS) is a novel optical technique that appears to be an excellent tool for assessing cerebral blood flow in a continuous and non-invasive manner at the bedside. We present new clinical validation of the DCS methodology by demonstrating strong agreement between DCS indices of relative cerebral blood flow and indices based on phase-encoded velocity mapping magnetic resonance imaging (VENC MRI) of relative blood flow in the jugular veins and superior vena cava. Data were acquired from 46 children with single ventricle cardiac lesions during a hypercapnia intervention. Significant increases in cerebral blood flow, measured both by DCS and by VENC MRI, as well as significant increases in oxyhemoglobin concentration, and total hemoglobin concentration, were observed during hypercapnia. Comparison of blood flow changes measured by VENC MRI in the jugular veins and by DCS revealed a strong linear relationship, R = 0.88, p < 0.001, slope = 0.91 +/- 0.07. Similar correlations were observed between DCS and VENC MRI in the superior vena cava, R = 0.77, slope = 0.99 +/- 0.12, p < 0.001. The relationship between VENC MRI in the aorta and DCS, a negative control, was weakly correlated, R = 0.46, slope = 1.77 +/- 0.45, p < 0.001. (C) 2012 Society of Photo-Optical Instrumentation Engineers (SPIE). [DOI: 10.1117/1.JBO.17.3.037007]
Subject: cerebral blood flow
diffuse optical spectroscopy
diffuse correlation spectroscopy
phase encoded velocity mapping magnetic resonance imaging
congenital heart defects
pediatrics
hypercapnia
Country: EUA
Editor: Spie-soc Photo-optical Instrumentation Engineers
Rights: aberto
Identifier DOI: 10.1117/1.JBO.17.3.037007
Date Issue: 2012
Appears in Collections:Artigos e Materiais de Revistas Científicas - Unicamp

Files in This Item:
File Description SizeFormat 
WOS000303697000028.pdf939.77 kBAdobe PDFView/Open


Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.