Please use this identifier to cite or link to this item: http://repositorio.unicamp.br/jspui/handle/REPOSIP/65048
Type: Artigo de periódico
Title: Effects of pH on the electrochemical behaviour of titanium alloys for implant applications
Author: Souza, MEP
Lima, L
Lima, CRP
Zavaglia, CAC
Freire, CMA
Abstract: The electrochemical behaviour of two commercial titanium alloys Ti-6Al-4 V (ASTM F136) and Ti-13Nb-13Zr (ASTM F1713) was investigated in Ringer physiological solution at two pH values (5.5 and 7.0). The corrosion properties were examined by using electrochemical techniques: Potentiodynamic anodic polarization, cyclic polarization and electrochemical impedance spectroscopy (EIS). The electrochemical corrosion properties of both alloys at different conditions were measured in terms of corrosion potential (E (corr)), corrosion current density (i (corr)) and passivation current density (i (pass)). Equivalent electrical circuits were used to modulate EIS data, in order to characterize alloys surface and better understanding the pH effect on the interface alloy/solution.
Country: Holanda
Editor: Springer
Rights: fechado
Identifier DOI: 10.1007/s10856-008-3623-y
Date Issue: 2009
Appears in Collections:Unicamp - Artigos e Outros Documentos

Files in This Item:
File Description SizeFormat 
WOS000263063300016.pdf315.51 kBAdobe PDFView/Open