Please use this identifier to cite or link to this item: http://repositorio.unicamp.br/jspui/handle/REPOSIP/58181
Type: Artigo de periódico
Title: Thermospray flame furnace atomic absorption spectrometry for determination of silver in biological materials
Author: Gerondi, F
Arruda, MAZ
Abstract: A method for silver determination without concentration steps is described using thermospray flame furnace atomic absorption spectrometry. Carrier type and flow rate, sample volume, flame conditions (acetylene and air flow rate), water flow rate in the nebulizer, metallic tube and type and concentration of the acid diluent of the analyte are the parameters evaluated in the optimization of the method. Using the optimized conditions, eleven elements are evaluated as concomitants. The limits of detection and quantification are 0.15 mu gL(-1) and 0.50 mu gL(-1), respectively. The linear range is from 0.50 mu gL(-1) to 40 mu gL(-1) and the accuracy of the method is obtained through two certified reference materials: MA-A-2 (fish flesh homogenate) and SRM 1643e (trace elements in water). (C) 2012 Elsevier B.V. All rights reserved.
Subject: TS-FF-AAS
Biological materials
Flow system
Silver
Country: Holanda
Editor: Elsevier Science Bv
Rights: fechado
Identifier DOI: 10.1016/j.talanta.2012.04.051
Date Issue: 2012