Please use this identifier to cite or link to this item: http://repositorio.unicamp.br/jspui/handle/REPOSIP/56715
Type: Artigo de periódico
Title: Microfluidic devices with integrated dual-capacitively coupled contactless conductivity detection to monitor binding events in real time
Author: Coltro, WKT
Neves, RD
Motheo, AD
da Silva, JAF
Carrilho, E
Abstract: This paper describes the fabrication of microfluidic devices with integrated dual capacitively coupled contactless conductivity detection ((CD)-D-4) for monitoring binding events. The avidin-biotin model was employed to show the feasibility of the (CD)-D-4 as a detection mode. The binding event was monitored using microfluidic channels fabricated in poly( dimethylsiloxane) (PDMS). Detection electrodes were fabricated in a three-electrode arrangement by sputtering a Ti/Au/Ti layer over a glass wafer, which were electrically insulated by a SiO2 layer. The dielectric layer over the sensing electrode area was chemically modified with biotin. After introducing avidin solutions on microchannels, binding events were detected on conductivity sensorgrams by applying a 300-kHz-sinusoidal waveform with amplitude of 2 V (peak-to-peak) to the exciting electrode. The electrical characterization of the modified electrode indicates that the conductivity signal is changed due to the alteration on the dielectric constant of the surface. The limit of detection for avidin was 1.5 mu g mL(-1), representing 23 nmol L-1. (C) 2013 Elsevier B.V. All rights reserved.
Subject: Contactless conductivity biosensor
Electrical transduction
Micro total analysis systems
Binding events
Country: Suíça
Editor: Elsevier Science Sa
Rights: fechado
Identifier DOI: 10.1016/j.snb.2013.10.114
Date Issue: 2014
Appears in Collections:Unicamp - Artigos e Outros Documentos

Files in This Item:
There are no files associated with this item.


Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.