Title: Analysis of adenosine by RP-HPLC method and its application to the study of adenosine kinase kinetics
Author: Marin, RM
Franchini, KG
Rocco, SA
Abstract: An RP-HPLC method for the analysis of adenosine (ADO) has been developed and validated. In the present study, we report an RP-HPLC-based method with modifications of mobile phase and shorter retention time that substantially improved the efficiency of ADO analysis. The HPLC separation of the ADO was achieved on a C18 column, using a mobile phase consisting of water, containing 7% v/v ACN, at a flow rate of 0.8 mL/min. The column effluent was monitored by UV detection at 260 nm. A linear response was achieved over the concentration range of 0.25-100.00 mu mol/L. The analytical method inter- and intra-run accuracy and precision were better than +/- 15%. The LOQwas 0.25 mu mol/L, with ADO detection in the range of 6.25 pmol per sample. The method has been applied to the study of adenosine kinase (AK) kinetics.
Subject: adenosine
adenosine kinase
kinetics
RP-HPLC
Country: Alemanha
Editor: Wiley-v C H Verlag Gmbh
Rights: fechado
Identifier DOI: 10.1002/jssc.200700194
Date Issue: 2007
Appears in Collections:Artigos e Materiais de Revistas Científicas - Unicamp