Please use this identifier to cite or link to this item: http://repositorio.unicamp.br/jspui/handle/REPOSIP/358530
Type: DISSERTAÇÃO DIGITAL
Degree Level: Mestrado
Title: Influência do campo de visão triangular convexo na expressão do artefato de alteração volumétrica em tomografia computadorizada de feixe cônico : Influence of the convex triangular field of view on the expression of volumetric alteration artefact in cone-beam computed tomography
Title Alternative: Influence of the convex triangular field of view on the expression of volumetric alteration artefact in cone-beam computed tomography
Author: Cascante-Sequeira, Deivi Antonio, 1983-
Advisor: Haiter Neto, Francisco, 1964-
Abstract: Resumo: Alguns aparelhos de tomografia computadorizada de feixe cônico (TCFC) oferecem a opção de um campo de visão (FOV) triangular convexo, além dos convencionais FOVs cilíndricos. A principal vantagem desse FOV triangular convexo é uma provável diminuição da dose de radiação recebida pelo paciente. Porém, é de grande importância saber se existe uma influência na alteração volumétrica decorrente da geometria da formação da imagem nesse FOV. O objetivo nesse estudo foi avaliar a influência do FOV triangular convexo na expressão da alteração volumétrica em comparação com FOVs cilíndricos de dimensões similares, em dois aparelhos de TCFC. Para isso, foram adquiridas imagens de TCFC de cinco cilindros de materiais de uso odontológico (amálgama, crômio-cobalto, guta-percha, titânio e zircônia) de volume físico conhecido. As aquisições foram realizadas sob condições padronizadas (90 kV, 5mA) usando dois formatos de FOVs (triangular convexo e cilíndrico) em dois aparelhos de TCFC (Veraviewepocs® R100 e Veraview® X800) sendo um material por vez, em quatro posições distintas, dentro de um fantoma de polimetilmetacrilato (PMMA). As imagens obtidas foram segmentadas de forma independente por dois radiologistas orais calibrados para a obtenção do volume (em mm³) utilizando o software ITK-SNAP V.3.8.0. A diferença entre os volumes aferidos e os volumes físicos correspondiam à expressão do artefato da alteração volumétrica. Esses valores resultantes foram submetidos ao teste de correlação intraclasse (ICC) e ao teste ANOVA para medidas repetidas com teste post-hoc de Tukey com nível de significância de 5%. O ICC mostrou concordâncias intra (0,999) e interavaliador (0,958) excelentes. O FOV influenciou no artefato de alteração volumétrica apenas no aparelho Veraview® X800 (p = 0,014). A alteração volumétrica no FOV triangular foi maior do que no FOV cilíndrico. No aparelho Veraviewepocs® R100, a alteração volumétrica foi semelhante tanto para o FOV triangular quanto para o cilíndrico. A comparação entre os dois FOVs triangulares mostrou que a alteração volumétrica no aparelho Veraview® X800 foi maior do que no aparelho Veraviewepocs® R100 (p = 0,000). O material influenciou na expressão do artefato de alteração volumétrica apenas no aparelho Veraviewepocs® R100 (p <0,001): a guta-percha apresentou maior alteração volumétrica, com subestimação, que os demais materiais: amálgama (p <0,001), titânio (p = 0,002), cromo-cobalto (p <0,001) e zircônio (p <0,001). Os demais materiais não diferiram entre si (p> 0,05), exceto o amálgama, que apresentou maior alteração volumétrica, sendo superestimado, do que o titânio (p = 0,042). Em relação à posição dos materiais de alta densidade dentro do FOV, não houve diferença na expressão da alteração volumétrica nos dois dispositivos (p> 0,05). Assim, concluímos que o FOV triangular aumentou a expressão do artefato de alteração volumétrica de materiais de alta densidade no aparelho Veraview® X800.

Abstract: Some Cone-Beam Computed Tomography (CBCT) devices offer the option of a convex triangular field of view (FOV), in addition to the conventional cylindrical FOVs. The main advantage of this triangular convex FOV is the probable decreasing of the patient radiation dose. However, it is important to know if there is an influence on the volumetric alteration due to the geometry of the image formation in this FOV. This study aimed to evaluate the influence of the convex triangular FOV on the expression of the volumetric alteration compared with cylindrical FOVs of similar dimensions. For this, CBCT images of five high-density dental material cylinders (amalgam, chromium-cobalt, gutta-percha, titanium, and zirconia) of known physical volume were acquired. The acquisitions were carried out under standardized conditions (90 kVp, 5mA) using two FOVs formats (convex triangular and cylindrical) on two CBCT devices (Veraviewepocs® R100 and Veraview® X800) one material at a time, in four different positions, inside of a polymethylmethacrylate phantom (PMMA). The images obtained were independently segmented by two calibrated oral radiologists to obtain de volume (in mm3) using the ITK-SNAP V.3.8.0 software. The difference between the measured volume and the physical volume corresponded to the volumetric alteration artifact expression. These resulting values were subjected to the intraclass correlation test (ICC) and the ANOVA test for repeated measures with Tukey's post-hoc test with a 5% significance level. The ICC showed excellent intra (0.999) and inter-evaluator (0.958) concordances The FOV influenced the volumetric alteration artifact only in the Veraview® X800 device (p=0.014). The volumetric alteration in the triangular FOV was greater than in the cylindrical FOV. In the Veraviewepocs® R100 device, the volumetric alteration was similar for both the triangular and the cylindrical FOVs. The comparison between both triangular FOVs showed that in the Veraview® X800 device, the volumetric alteration was greater than in the Veraviewepocs® R100 device (p=0.000). The material influenced the expression of the volumetric alteration artifact only in the Veraviewepocs® R100 device (p<0.001). The Tukey's post-hoc test showed that gutta-percha presented higher volumetric alteration, being underestimated, than the other materials: amalgam (p <0.001), titanium (p=0.002), chromium-cobalt (p<0.001), and zirconium (p0.05), except for amalgam, which presented higher volumetric alteration, being overestimated, than titanium (p=0.042). Regarding the position of high-density materials within the FOV, there was no difference on the expression of volumetric alteration in both devices (p>0.05). Thus, we conclude that the triangular FOV increased the volumetric alteration artifact expression of high-density materials in the Veraview® X800 device.
Subject: Artefatos
Tomografia computadorizada de feixe cônico
Materiais dentários
Language: Multilíngua
Editor: [s.n.]
Citation: CASCANTE-SEQUEIRA, Deivi Antonio. Influência do campo de visão triangular convexo na expressão do artefato de alteração volumétrica em tomografia computadorizada de feixe cônico : Influence of the convex triangular field of view on the expression of volumetric alteration artefact in cone-beam computed tomography . 2021. 1 recurso online ( 50 p.) Dissertação (mestrado) - Universidade Estadual de Campinas, Faculdade de Odontologia de Piracicaba, Piracicaba, SP.
Date Issue: 2021