Title: Direct measurement of clinical mammographic X-ray spectra using a CdTe spectrometer
Author: Santos, J. C.
Tomal, A.
Furquim, T. A.
Fausto, A. M. F.
Nogueira, M. S.
Costa, P. R.
Abstract: Purpose: To introduce and evaluate a method developed for the direct measurement of mammographic x-ray spectra using a CdTe spectrometer. The assembly of a positioning system and the design of a simple and customized alignment device for this application is described. Methods: A positioning system was developed to easily and accurately locate the CdTe detector in the x-ray beam. Additionally, an alignment device to line up the detector with the central axis of the radiation beam was designed. Direct x-ray spectra measurements were performed in two different clinical mammography units and the measured x-ray spectra were compared with computer-generated spectra. In addition, the spectrometer misalignment effect was evaluated by comparing the measured spectra when this device is aligned relatively to when it is misaligned. Results: The positioning and alignment of the spectrometer have allowed the measurements of direct mammographic x-ray spectra in agreement with computer-generated spectra. The most accurate x-ray spectral shape, related with the minimal HVL value, and high photon fluence for measured spectra was found with the spectrometer aligned according to the proposed method. The HVL values derived from both simulated and measured x-ray spectra differ at most 1.3 and 4.5% for two mammography devices evaluated in this study. Conclusion: The experimental method developed in this work allows simple positioning and alignment of a spectrometer for x-ray spectra measurements given the geometrical constraints and maintenance of the original configurations of mammography machines.
Subject: Mamas - Radiografia
Breast - Radiography
Country: Estados Unidos
Editor: Wiley-Blackwell Publishing
Rights: fechado
Identifier DOI: 10.1002/mp.12287
Address: https://aapm.onlinelibrary.wiley.com/doi/full/10.1002/mp.12287
Date Issue: 2017
Appears in Collections:IFGW - Artigos e Outros Documentos

Files in This Item:
File SizeFormat 
000405103800014.pdf745.12 kBAdobe PDFView/Open


Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.