Please use this identifier to cite or link to this item: http://repositorio.unicamp.br/jspui/handle/REPOSIP/340391
Type: Artigo
Title: Impact of movement and motion-artefact correction on image quality and interpretability in CBCT units with aligned and lateral-offset detectors
Author: Santaella, Gustavo Machado
Wenzel, Ann
Haiter-Neto, Francisco
Rosalen, Pedro Luiz
Spin-Neto, Rubens
Abstract: To evaluate the impact of movement and motion-artefact correction systems on CBCT image quality and interpretability of simulated diagnostic tasks for aligned and lateral-offset detectors. Methods: A human skull simulating three diagnostic tasks (implant planning in the anterior maxilla, implant planning in the left-side-mandible and mandibular molar furcation assessment in the right-side-mandible) was mounted on a robot performing six movement types. Four CBCT units were used: Cranex 3Dx (CRA), Ortophos SL (ORT), Promax 3D Mid (PRO), and X1. Protocols were tested with aligned (CRA, ORT, PRO, and X1) and lateral-offset (CRA and PRO) detectors and two motion-artefact correction systems (PRO and X1). Movements were performed at one moment-in-time (t(1)), for units with an aligned detector, and three moments-in-time (t(1)-first-half of the acquisition, t(2)-second-half, t(3)-both) for the units with a lateral-offset detector. 98 volumes were acquired. Images were scored by three observers, blinded to the unit and presence of movement, for motion-related stripe artefacts, overall unsharpness, and interpretability. Fleiss' kappa was used to assess interobserver agreement. Results: Interobserver agreement was substantial for all parameters (0.66-0.68). For aligned detectors, in all diagnostic tasks a motion-artefact correction system influenced image interpretability. For lateral-offset detectors, the interpretability varied according to the unit and moment-in-time, in which the movement was performed. PRO motion-artefact correction system was less effective for the offset detector than its aligned counterpart. Conclusion: Motion-artefact correction systems enhanced image quality and interpretability for units with aligned detectors but were less effective for those with lateral-offset detectors
Subject: Tomografia computadorizada de feixe cônico
Country: Reino Unido
Editor: British Institute of Radiology
Rights: Fechado
Identifier DOI: 10.1259/dmfr.20190240
Address: https://www.birpublications.org/doi/full/10.1259/dmfr.20190240
Date Issue: 2019
Appears in Collections:FOP - Artigos e Outros Documentos

Files in This Item:
File Description SizeFormat 
000502611400005.pdf1.18 MBAdobe PDFView/Open