Title: Beam hardening correction tool improves the diagnosis of incipient caries lesions in micro-CT images
Author: Rovaris, Karla
Melo, Saulo Leonardo Sousa
Queiroz, Polyane Mazucatto
Loch, Carolina
Schwass, Donald R.
Haiter-Neto, Francisco
Abstract: To evaluate the effect of Beam Hardening Correction tool (BHC) on Micro-CT images used in the diagnosis of incipient caries lesions. Methods: Six posterior teeth, with 6 sound proximal surfaces and 6 natural white spot lesions, were scanned using a SkyScan1174 device. Each tooth was imaged 16 times in various acquisition parameters, and reconstructed 3 times applying different levels of BHC (0, 15 and 30%). The 288 datasets were analyzed by three examiners. Gold-standard was obtained by microscopic analyses under polarized light. Area under the ROC curve was obtained for each protocol and the BHC parameter was isolated to individually observe the influence of such parameters of reconstruction. Results: Microtomographic images were accurate for the diagnosis of incipient caries lesions even when the BHC tool was not applied. Accuracy was perfect in 50% of the protocols without BHC. When adding the artefact correction, the number of protocols that reached perfect accuracy increased to 81.25%. No difference was found between the protocols that used 15 or 30% BHC. Statistically significant differences were found only in four comparisons, and on those the images that had BHC were more accurate. Two non-BHC protocols reached accuracy scores of 0.653 and 0.881, and were improved to perfect accuracy with BHC application. Conclusion: The use of the BHC tool should be encouraged when using Micro-CT to detect incipient caries lesions, especially when low-quality parameters aiming less acquisition and processing time are used
Subject: Microtomografia por raio-X
Cárie dentária
Country: Reino Unido
Editor: British Institute of Radiology
Rights: Fechado
Identifier DOI: 10.1259/dmfr.20180123
Address: https://www.birpublications.org/doi/full/10.1259/dmfr.20180123
Date Issue: 2019
Appears in Collections:FOP - Artigos e Outros Documentos

Files in This Item:
File Description SizeFormat 
000453231900008.pdf267.15 kBAdobe PDFView/Open


Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.