Title: Evaluation of a metal artefact reduction tool on different positions of a metal object in the FOV
Author: Queiroz, Polyane M.
Santaella, Gustavo M.
da Paz, Thais D. J.
Freitas, Deborah Q.
Abstract: To evaluate the action of a metal artefact reduction (MAR) tool when artefact-generator metal object is at different positions in the field of view (FOV). A cylindrical utility wax phantom, with a metal alloy sample inside, was made. The phantom was positioned centrally and peripherally in the FOV for image acquisition, with and without the MAR tool activation. The standard deviation values (image noise levels) from areas around the metal sample and the control area were obtained. The numbers were compared by Student's t-test (alpha=0.05). When the tool was activated, a significant difference of image noise was observed for central and peripheral positioning, for both control area (p=0.0012) and metal area (p= 0.03), and a smaller level of noise was observed for images with phantoms in central positioning. A decrease in image noise with the tool activated was found only in phantoms with the metal object positioned centrally in the FOV. For the MAR tool to be effective, the artefact-generator object needs to be in the central region of the FOV
Subject: Artefatos
Tomografia computadorizada de feixe cônico
Country: Reino Unido
Editor: British Institute of Radiology
Citation: Dentomaxillofacial Radiology. British Inst Radiology , v. 46, p. , 2017.
Rights: fechado
Identifier DOI: 10.1259/dmfr.20160366
Address: https://www.birpublications.org/doi/full/10.1259/dmfr.20160366
Date Issue: 2017
Appears in Collections:FOP - Artigos e Outros Documentos

Files in This Item:
File SizeFormat 
000395933400002.pdf272.53 kBAdobe PDFView/Open


Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.