Please use this identifier to cite or link to this item: http://repositorio.unicamp.br/jspui/handle/REPOSIP/202055
Type: Artigo de periódico
Title: Evaluation Of Characteristic-to-total Spectrum Ratio: Comparison Between Experimental And A Semi-empirical Model.
Author: Lopez Gonzales, A H
Tomal, A
Costa, P R
Abstract: Primary X-ray spectra were measured in the range of 80-150kV in order to validate a computer program based on a semiempirical model. The ratio between the characteristic and total air Kerma was considered to compare computed results and experimental data. Results show that the experimental spectra have higher first HVL and mean energy than the calculated ones. The ratios between the characteristic and total air Kerma for calculated spectra are in good agreement with experimental results for all filtrations used.
Subject: Cdte Detector.
Characteristic-to-total Spectrum Ratio
X-ray Spectrum
Rights: fechado
Identifier DOI: 10.1016/j.apradiso.2015.01.011
Address: http://www.ncbi.nlm.nih.gov/pubmed/25600507
Date Issue: 2015
Appears in Collections:Artigos e Materiais de Revistas Científicas - Unicamp

Files in This Item:
File SizeFormat 
pmed_25600507.pdf772 kBAdobe PDFView/Open


Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.