Please use this identifier to cite or link to this item: http://repositorio.unicamp.br/jspui/handle/REPOSIP/194967
Type: Artigo de periódico
Title: A Portable Fiber-optic Chemical Device For The Quantitative Determination Of Carbon Monoxide From Automobile Exhaust Emissions.
Author: Matias, F A
Tubino, M
Abstract: A colorimetric method for the quantitative determination of CO by diffuse reflectance is described. This method is based on the reduction by CO of Mo (VI) from the indicator reagent molybdosilicic acid (H8Si[Mo2O7]6). The reduction yielded a change of color from clear yellow to dark green on white disk filter chart paper wetted with reagent indicator solution. The gaseous mixture containing CO was forced to pass through this chart paper, initiating the reaction. The intensity of the color produced, measured by diffuse reflectance, was proportional to the CO concentration present in exhaust gases in the range from 0.02 to 12% volume/volume (v/v). A 650-nm light-emitting diode was used as a light source. A two-fiber-optic system carried the light from the source to the detection system, which was composed of a photodiode, an amplification circuit, and a digital display. The method was applied with success in field measurements for automobiles in the Otto cycle. In a previous paper, this method was used for the quantitative determination of exhaust emissions from diesel-fueled vehicles.
Subject: Carbon Monoxide
Environmental Monitoring
Fiber Optic Technology
Optical Fibers
Vehicle Emissions
Rights: fechado
Identifier DOI: 
Address: http://www.ncbi.nlm.nih.gov/pubmed/15658214
Date Issue: 2001
Appears in Collections:Artigos e Materiais de Revistas Científicas - Unicamp

Files in This Item:
There are no files associated with this item.


Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.