Title: Design, Manufacturing And Tests Of An Implantable Centrifugal Blood Pump
Author: Bock E.
Antunes P.
Uebelhart B.
Leao T.
Fonseca J.
Cavalheiro A.
Santos Filho D.
Cardoso J.R.
Utiyama B.
Leme J.
Silva C.
Andrade A.
Arruda C.
Abstract: An implantable centrifugal blood pump was developed for long-term ventricular assistance in cardiac patients. In vitro tests were performed, as wear evaluation, performance tests and hemolysis tests in human blood. Numerical computational simulations were performed during design process in order to predict its best geometry. Wear evaluations helped to select the best materials for double pivot bearing system proposed to achieve longer durability. Performance tests pointed the best impeller geometry. The implantable centrifugal blood pump was compared with other blood pumps founded in literature. The proposed implantable centrifugal blood pump showed the best performance. But, its results showed a strong descendant curve in high flow. Other prototype was manufactured with a different inlet port angle to overcome this problem. The normalized index of hemolysis (NIH) measured 0.0054 mg/100L that can be considered excellent since it is close to the minimum found in literature (between 0.004 g/ 100L e 0.02 g/ 100L). The authors' expectation is that this pump will become a promising Technological Innovation for Sustainability. © 2011 IFIP International Federation for Information Processing.
Editor: 
Rights: fechado
Identifier DOI: 10.1007/978-3-642-19170-1_45
Address: http://www.scopus.com/inward/record.url?eid=2-s2.0-79952222294&partnerID=40&md5=57834121ba8d9c70be68096973ddc5c9
Date Issue: 2011
Appears in Collections:Unicamp - Artigos e Outros Documentos

Files in This Item:
File SizeFormat 
2-s2.0-79952222294.pdf19.42 MBAdobe PDFView/Open


Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.