Please use this identifier to cite or link to this item: http://repositorio.unicamp.br/jspui/handle/REPOSIP/106178
Type: Artigo de periódico
Title: Construction Of A Long-pathlength Flow Cell For Spectrophotometric Measurements [construção De Uma Cela De Fluxo De Longo Caminho óptico Para Medidas Espectrofotométricas]
Author: Melchert W.R.
Rocha F.R.P.
Raimundo Jr. I.M.
Reis B.F.
Abstract: A simple and low-cost flow cell with 30 cm optical path for spectrophotometric measurements is described. It presents desirable characteristics such as low attenuation of the radiation beam and internal volume (75 μL) comparable to that of a 1-cm conventional cell (80 μL). Despite the increase in optical path, the effect on sample dispersion was also similar to that attained in the commercial cell. The performance of the cell was assessed by the determination of phosphate based on the molybdenum blue method, yielding a linear response range between 0.05 and 0.8 mg L-1 phosphorus (r=0.999). The increase in sensitivity (30.4-fold) in comparison with that obtained with a conventional 1-cm flow cell agreed with that estimated by the Lambert-Beer law.
Editor: 
Rights: aberto
Identifier DOI: 10.1590/S0100-40422008000200042
Address: http://www.scopus.com/inward/record.url?eid=2-s2.0-42149166781&partnerID=40&md5=f252a2f38ff7b65577dd3dea34fd0d94
Date Issue: 2008
Appears in Collections:Unicamp - Artigos e Outros Documentos

Files in This Item:
File Description SizeFormat 
2-s2.0-42149166781.pdf145.63 kBAdobe PDFView/Open


Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.